C9830PDE Near infrared fluorescence imager

c9830 product photo

This camera system is used to observe invisible near-infrared images, which makes it possible to observe the epidermic area or distribution of fluorescence reagent inside the tissue.
510(k) Number: K110480

Available for sales only in China.

Features

Easy of handling to observe the area of interest by handy-type camera head

■ Small and compact sized handy-type camera unit

c9830 product feature


■ Real time observation is possible by holding the camera unit closer to the area of interest.


■ Operation is possible by either manually or remote controller

c9830 product feature

Go to top

ICG fluorescence method

Observation by ICG fluorescence

When ICG is administered as a bolus intravenously, the PDE is able to visualize the ICG fluorescence allowing a qualitative assessment of blood flow and tissue perfusion.


Fluorescence characteristic of ICG

After bonding with plasma protein in the blood, ICG will become excited with near infrared light and fluoresce at a slightly longer near infrared wavelength. The PDE's special sensor and filters will see this fluorescence clearly through a range of human soft tissue.

c9830 product feature

Go to top

Highly prospective observation method

ICG fluorescence method is highly expected for use in the field of breast cancer surgery, plastic surgery, digestive surgery and others as the innovative less-invasive method.

c9830 product feature


a) Identification of hepatic segments

Case: Possible to secure the Glisson’s sheath of the dominant region from the porta hepatis

Identify and secure the Glission’s sheath and then insert a clamp before intravenously injecting ICG for the PDE observation.

c9830 measurement example


Case: Impossible to secure the Glission’s sheath of the dominant region from the porta hepatis

Use intraoperative ultrasonography to pucture the dominant Glission’s sheath and inject the ICG. Clamp the portal vein and perform the PDE observation.

c9830 measurement example


b) Intraoperative cholangiography

Case: Intraoperative fluorescent cholangiography: Intraductal injection method

While observing the fluorescence with PDE, inject ICG into the bile duct through and indwelling imaging catheter.

c9830 measurement example


Case: Intraoperative fluorescent cholangiography: Intravenous injection method

Intravenously inject ICG one hour before the start of surgery, and perform imaging with PDE.

c9830 measurement example


c) Esophageal cancer surgery: Evaluation of blood flow in reconstructed organs

Possible to observe perfusion level of the anastomosis by fluorescence image.

c9830 measurement example

Go to top

Configuration

c9830 system configuration

Go to top

Specifications

Type number  C9830 
Power supply voltage  AC 100 V to AC 240 V 
Power frequency  50 Hz / 60 Hz 
Power consumption  Max. 60 VA 
Ambient operating temperature  +10 ℃ to +30 ℃ 
Ambient operating humidity  20 % to 70 % (with no condensation) 
Ambient storage temperature  0 ℃~+40 ℃ 
Ambient storage humidity  20 % to 70 % (with no condensation) 
Dimensions (not including projections)
Weight (not including cables and accessories) 
Camera unit: Approx. 80 mm (W)×181 mm (D)×80 mm (H), Approx. 0.5 kg
Controller: Approx. 322 mm (W)×283 mm (D)×55 mm (H), Approx. 2.8 kg 
Controller function  Noise reduction, Contrast enhancement 
Output signal  Black and White (EIA) 
Video output  2 BNC 

Go to top

Related information

Go to top

Go to top

Disclaimer
 [66KB/PDF]