Spectrometers Spectrometers

Learn about spectrometer specifications | Spectrometers

Spectral range

Spectral range refers to the range of wavelengths that a spectrometer can detect. The spectral range of a spectrometer is basically determined by the groove density of the grating and the number of pixels in the horizontal (wavelength-axis) direction of the sensor.

When the groove density of the grating differs

When the number of pixels is the same, a lower (coarser) grating groove density results in a wider spectral range but lower spectral resolution (see Figure 1).

Conversely, a higher (finer) grating groove density results in a narrower spectral range but higher spectral resolution (see Figure 2).

Figure 1: When the groove density of the grating is low (coarse)

Figure 2: When the groove density of the grating is high (fine)

When the number of sensor pixels differs

When the grating groove density is the same, a larger number of sensor pixels results in a wider spectral range (see Figure 3), while a smaller number of pixels results in a narrower spectral range (see Figure 4).

*The pixel size is assumed to be the same.

Figure 3: When the number of sensor pixels is large

Figure 4: When the number of sensor pixels is small

Spectral resolution

Spectral resolution is an indicator of how finely a spectrometer can distinguish between wavelengths when detecting light. For example, a spectrometer with a spectral resolution of 1 nm can distinguish between 500 nm and 501 nm, but a spectrometer with a spectral resolution of 5 nm cannot distinguish between them. In the case of a polychromator, spectral resolution is determined by a combination of factors, including the entrance slit width, the grating groove density, and the sensor pixel size. (Strictly speaking, the optical design inside the spectrometer and the performance of the optical components used also have an effect.)

When the groove density of the grating differs

This point is the same that was explained in the "Spectral range.”

When the number of sensor pixels is the same, a lower (coarser) grating groove density results in a wider spectral range but lower spectral resolution (see Figure 5). Conversely, a higher (finer) grating groove density results in a narrower spectral range but higher spectral resolution (see Figure 6).

Figure 5: When the groove density of the grating is low (coarse)

Figure 6: When the groove density of the grating is high (fine)

When the width of the entrance slit differs

The entrance slit serves to limit the spatial dispersion of light entering the spectrometer. When the optical system inside the spectrometer forms an image at unity magnification, a narrower entrance slit width results in higher resolution, while a wider entrance slit width results in lower resolution. However, even if the entrance slit is narrower than the pixel size, a resolution finer than the pixel size cannot be achieved.

When the entrance slit width is narrower than the pixel size (Figure 7), only light of a specific wavelength enters a given pixel. In contrast, when the entrance slit width is wider than the pixel size (Figure 8), light of a given wavelength enters not only one pixel but also the adjacent pixels, resulting in lower resolution. 

Figure 7: When the entrance slit width is narrower than the pixel size

Figure 8: When the entrance slit width is wider than the pixel size

When the pixel size differs

When the sensor pixel size is large (Figure 9), multiple wavelengths enter the same pixel even if the grating finely separates them , resulting in lower spectral resolution.

When the sensor pixel size is small (Figure 10), the light separated by the grating enters a separate pixel for each wavelength, allowing the wavelengths to be resolved.

Figure 9: When the sensor pixel size is large

Figure 10: When the sensor pixel size is small

Recommended products for high spectral resolution

Spectral range: 200 nm to 900 nm 
Spectral resolution: ≤0.9 nm

Spectral range: 790 nm to 1050 nm 
Spectral resolution: ≤0.4 nm

Spectral range: 200 nm to 800 nm 
Spectral resolution: ≤1.0 nm

Dynamic range

Dynamic range is the range between the maximum and minimum signal levels that a spectrometer can measure and is defined as the full well capacity divided by the readout noise. The higher the dynamic range, the better a spectrometer can measure both weak and strong signals at the same time. With a low dynamic range, configuring the spectrometer to detect weak signals causes strong signals to saturate, while configuring it to detect a strong signal makes weak signals undetectable.

Applications that require a high dynamic range include measuring strong light, such as laser light or plasma emission, and measuring extremely weak light, such as band-edge emission from luminescent materials.

Measurement example with the high dynamic range spectrometer OPAL-Luxe

レーザ計測

With conventional spectrometers, when measuring intense light such as laser light, increasing the exposure time causes the detector to saturate due to the strong laser signal. This makes it difficult to simultaneously measure low-level light at other wavelengths, oftern requiring the laser light to be blocked using a filter or similar component. The OPAL-Luxe delivers high S/N performance even for low-level light, making it well suited for applications such as plasma emission measurement, where strong and weak emissions are coexist.

ガリウムナイトライド計測

This is an example of a photoluminescence (PL) measurement of the compound semiconductor gallium nitride (GaN) excited with a helium-cadmium laser (wavelength: 325 nm). The OPAL-Luxe successfully measured band-edge emission and a fluorescence spectra with a high S/N, which could not be detected with our previous model (PMA-12).

Recommended products for high dynamic range

Dynamic range: 2500000 : 1

Sensors

CCD sensors and CMOS sensors are commonly used as detectors in spectrometers (polychromators) for the visible region. Some CCD sensors are available with TE-cooled type and tend to exhibit lower dark current. The greatest advantage of CCD sensors is that they support vertical binning. Binning is a method in which multiple pixels are combined and treated as a single pixel for readout. Among the various binning methods, vertical binning is a method in which signals from pixels along the vertical direction of the sensor are added together and read out. By combining pixels in the vertical direction and treating them as a single pixel for readout, the signal level increases in proportion to the number of pixels added together, while the noise generated during readout occurs only once, resulting in improved S/N.

The CMOS image sensors used in mini-spectrometers are linear type, and are available in passive-pixel and active-pixel types.

The passive-pixel type has the advantage of high output uniformity because, like CCD sensors, it uses a single amplifier. It also has a full-well capacity, making it widely used in spectroscopic applications. The active-pixel type has a readout amplifier for each pixel; while its output uniformity and full-well capacity are inferior to those of the passive-pixel type, it offers the advantages of high sensitivity and fast readout speed.

Figure 11: Schematic diagram of a CCD sensor

パッシブピクセル型CMOSセンサの模式図

Figure 12: Schematic diagram of a passive-pixel CMOS sensor

アクティブピクセル型CMOSセンサの模式図

Figure 13: Schematic diagram of an active-pixel CMOS sensor

Sensors CCD Passive-pixel CMOS sensor Active-pixel CMOS sensor
Noise variation between pixels ◎ ◎ △
Full-well capacity ◎ ◎ △
Dark current noise ◎ 〇 △
Repetition rate △ △ ◎
Miniaturization and cost reduction △ ◎ ◎

Recommended products

With a CCD sensor​

Spectral range: 200 nm to 900 nm 
Spectral resolution: ≤0.9 nm

Spectral range: 200 nm to 800 nm 
Spectral resolution: ≤4.0 nm

With a passive-pixel CMOS sensor​

Spectral range: 340 nm to 780 nm 
Spectral resolution: ≤12 nm

With an active-pixel CMOS sensor​

Spectral range: 340 nm to 830 nm 
Spectral resolution: ≤2.3 nm

Spectral range: 190 nm to 440 nm 
Spectral resolution: ≤5.5 nm

Exposure time

Exposure time refers to the period during which the CCD sensor or CMOS sensor incorporated in a spectrometer accumulates signal. The longer the exposure time, the more photons enter the sensor, allowing low-level light signals to be captured. On the other hand, a longer exposure time also increases the time required for measurement, reducing the number of measurements that can be performed per unit of time.

For example, with an exposure time of 100 ms, 10 measurements can be performed per second, whereas with an exposure time of 1 s, only one measurement can be performed per second. 

 

*Strictly speaking, the time required to read out the sensor must also be taken into account when calculating the repetition rate. 

Exposure time  Longer exposure time Shorter exposure time
Sensitivity High Low
Dark current noise High Low
Repetition rate Slow Fast

Repetition rate

Repetition rate refers to the speed at which a spectrometer can acquire and process data during measurement. The repetition rate of a monochromator mainly depends on the rotation speed of the grating, while the repetition rate of a polychromator mainly depends on the readout speed of the sensor.  The faster the repetition rate, the faster changes in the measured light can be captured.

In general, monochromators tend to have a slower repetition rate because the grating angle must be rotated, while polychromators tend to have a faster repetition rate because multiple wavelengths can be measured simultaneously without moving the grating. However, in polychromators, the faster the CCD or CMOS readout speed, the higher the readout noise tends to be, so there is a trade-off between measurement speed and noise. A faster repetition rate allows more accumulation (averaging) within the same measurement time, which can improve the S/N of the measured data.

Spectrometer Monochromator Polychromator
Repetition rate Slow Fast

Sensors CCD CMOS
Repetition rate Slow Fast

*Among polychromators, those with a CCD sensor generally prioritize sensitivity over repetition rate, while those with a CMOS sensor generally prioritize speed.

Dark Current

Dark current is noise that generated by heat within the sensor. This noise can be reduced by cooling the sensor, but the additional components required for cooling have the disadvantage of increasing the size of the housing. In addition, since dark current is noise that accumulates as the exposure time increases, it is recommended to use a spectrometer with a cooling function when measuring extremely weak signals that cannot be detected without a long exposure time.

How to choose a spectrometer

Hamamatsu Photonics offers a lineup of various spectrometer types, including high-performance benchtop spectrometers, mini-spectrometers integrating an optical system, image sensor, and readout circuit into a compact housing, and extremely small, fingertip-sized spectrometer heads realized using MEMS and image sensor technologies.

This page provides tips for selecting the spectrometer that best suits your application from Hamamatsu's diverse spectrometer lineup. You can compare product specifications by spectral range, spectral resolution, and housing size.

Contact us for more information.