LIGHTNINGCURE LC-L1V5

L14310 / L14311 / C14052 series

Compact body easily installs into narrow confined spaces.
We came up with a unit that drives 4 heads but is small enough to fit in the palm of your hand. This unit can also be freely placed standing or horizontal in just a tiny space, so it needs no special layout. Our unique feedback function minimizes drift during initial light emission period and constantly maintains fluctuations in light output within 5 %.

Specifications

LED head unit L14310 series

Parameter Standard type Unit
Irradiation area
φ3 mm type
Irradiation area
φ6 mm type
Irradiation area
φ8 mm type
Irradiation area
φ12 mm type
L14310-□10
(*1) (*2)
L14310-□15
(*1) (*2)
L14310-□20
(*1) (*2)
L14310-□00
(*1) (*2)
Mounted condenser lens E11923-010 E11923-015 E11923-020
Focal length 10 15 20 10 mm
Irradiation area (*3) Approx. φ3 Approx. φ6 Approx. φ8 Approx. φ10 mm
Wavelength 365 / 385 / 405 nm
UV irradiance (*4) 14000 6500 4000 1000 mW/cm2
LED design life (*5) 20000 h
Cooling method Not required
Operating temperature range +5 ℃ to +40 ℃
Storage temperature range -10 ℃ to +60 ℃ (no freezing)
Operating humidity range 20 % to 80 % (no condensation)
Storage humidity range Below 80 % (no condensation)
Applicable standards EMC standards IEC 61326-1
Emission limits: CISPR 11 Group 1 Class A
Immunity requirements: Table 2
Safety standards IEC/EN 61010-1/A1

*1: This type number indicates a combination of L14310-□00 LED head unit and E11923 series condenser lens.
     The focal length and irradiation area can be changed by replacing the E11923 series condenser lens.

*2: The square symbol □ in each type number indicates a suffix that represents the wavelength.
     1: 365 nm, 2: 385 nm, 4: 405 nm.

*3: Irradiation area measured at the focal length.

*4: Maximum UV irradiance within the irradiated area measured at the focal length.

*5: Average time required for UV irradiance to reach 70 % of the initial value at an operating temperature of 25 ℃.

LED head unit L14311 series

Parameter Right-angle type Linear beam type  
Mid focal length type Long focal length type Wide range type  
L14311-□03 (*2) L14311-□05 (*2) L14311-□02 (*2) L14311-□04 (*2)
Mounted condenser lens
Focal length 10 20 15 10 mm
Irradiation area (*3) Approx. φ6 Approx. φ7 Approx. φ20 x 6 Approx. φ12 x 4 mm
Wavelength 365 / 385 / 405 nm
UV irradiance (*4) 4000 3000 1200 3000 mW/cm2
LED design life (*5) 20000 h
Cooling method Not required
Operating temperature range +5 ℃ to +40 ℃
Storage temperature range -10 ℃ to +60 ℃ (no freezing)
Operating humidity range 20 % to 80 % (no condensation)
Storage humidity range Below 80 % (no condensation)
Applicable standards EMC standards IEC 61326-1
Emission limits: CISPR 11 Group 1 Class A
Immunity requirements: Table 2
Safety standards IEC/EN 61010-1/A1

*2: The square symbol □ in each type number indicates a suffix that represents the wavelength.
     1: 365 nm, 2: 385 nm, 4: 405 nm.

*3: Irradiation area measured at the focal length.

*4: Maximum UV irradiance within the irradiated area measured at the focal length.

*5: Average time required for UV irradiance to reach 70 % of the initial value at an operating temperature of 25 ℃.

Condenser lens E11923 series

Parameter E11923-010 E11923-015 E11923-020 Unit
Focal length 10 15 20 mm
Irradiation area (*6) Approx. φ3 Approx. φ6 Approx. φ8 mm
Operating temperature range +5 ℃ to +40 ℃
Storage temperature range -10 ℃ to +60 ℃ (no freezing)
Operating humidity range 20 % to 80 % (no condensation)
Storage humidity range Below 80 % (no condensation)

*6: Irradiation area measured at the focal length.

Controller C14052 series

Parameter C14052-0-□□ (*7) C14052-1-□□ (*7) C14052-2-□□ (*7) Unit
Input voltage (AC) 100 V to 240 V, single phase 50 Hz / 60 Hz (*8)
Power consumption (*9) Max. 25 W
Cooling method Not required
Operating temperature range +5 ℃ to +40 ℃
Storage temperature range -10 ℃ to +60 ℃ (no freezing)
Operating humidity range 20 % to 80 % (no condensation)
Storage humidity range Below 80 % (no condensation)
External control Irradiation control, irradiation signal, various error signals
Communication control N/A Control via USB Control via RS-232C
Irradiation control, irradiation signal,
accumulation / display / reset of irradiation time,
program setting for auto irradiation, various error signals
Applicable standards EMC standards IEC 61326-1
Emission limits: CISPR 11 Group 1 Class A
Immunity requirements: Table 2
Safety standards IEC/EN 61010-1/A1

*7: The square symbol □ in each type number indicates a suffix that represents the specifications of the AC adapter.
     ・When no AC adapter supplied, no suffix is added. The user needs to prepare an AC adapter that outputs 12 V DC and 24 W or more.
         In that case, EMC and safety testing should be performed by the user.
     ・When an AC adapter supplied, a suffix is added. A1: For Japan, A2: For U.S., A3: For Europe, A4: For China, A5: UK, A7: Thailand
     In the case of C14052-0, the "-0" will be omitted and so the type number will be C14052.

*8: Input voltage to the controller should be 12 V ±0.5 V DC.

*9: Power consumption when 4 heads are operated.

Characteristics

Irradiance distributions (Typ. 365 nm)

Standard type: Irradiation area φ3 mm type

Standard type: Irradiation area φ6 mm type

Standard type: Irradiation area φ8 mm type

Standard type: Irradiation area φ12 mm type

NOTE: The center of the external dimensions of the LED head unit may not coincide with the center of the irradiated light. Adjust the installation position as needed.

    The shape of the irradiated light varies depending on the irradiation distance, for example, becomes close to a square.

    There are individual differences in the irradiance distribution between LED head units due to variations in the emission intensity of LED elements.

Demo unit loaning

Demo unit is available.

Please click below and contact us for the details.

Contact us for more information.